Key result
A&D UA-651 BP monitor passes European Society of Hypertension validation with wide-range cuff.
Why the study?
The accuracy of the A&D BP UA-651 device with a wide-range cuff for home blood pressure measurement needed validation according to the European Society of Hypertension International Protocol revision 2010.
Observational (n=33)
Device validation in limited cohort supports cuff options for home BP; leaves open broader accuracy and outcome data.
The aim of this study was to determine the accuracy of the A&D BP UA-651 device coupled to a wide-range cuff for home blood pressure (BP) measurement according to the International Protocol of the European Society of Hypertension. The device was evaluated in 33 patients. The mean age of the patients was 56.5±15.1 years. The mean systolic BP was 144.3±23.8 mmHg (range 88 : 196), the mean diastolic BP was 87.5±15.8 mmHg (range 38 : 132), and the mean arm circumference was 29.0±3.4 cm (range 22 : 36). The protocol requirements were followed precisely. The device passed all requirements, fulfilling the standards of the protocol. On average, the device overestimated the systolic BP by 0.7±3.4 mmHg and underestimated the diastolic BP by 0.8±3.6 mmHg. The measurement error was unrelated to the patient's arm circumference. These data show that the A&D BP UA-651 device coupled to a wide-range cuff fulfilled the requirements for validation by the International Protocol over a wide range of arm circumferences and can be recommended for clinical use in the adult population.
No takes yet. Share an insight, caveat, or question.
Benetti et al. (2015) conducted an observational in Blood pressure measurement (n=33). A&D BP UA-651 device with a wide-range cuff vs. Standard protocol reference was evaluated on Device accuracy according to the European Society of Hypertension International Protocol (Mean difference of 0.7±3.4 mmHg for systolic BP and -0.8±3.6 mmHg for diastolic BP). The A&D BP UA-651 device with a wide-range cuff fulfilled the European Society of Hypertension International Protocol requirements, overestimating systolic BP by 0.7±3.4 mmHg and underestimating diastolic BP by 0.8±3.6 mmHg.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: