Key result
Higher preimplantation BNP linked to ~119% increased risk of appropriate ICD therapies.
Why the study?
The ability of preimplantation B-type natriuretic peptide and C reactive protein concentrations to predict future appropriate ICD therapies was uncertain.
Cohort (n=345)
No
Answers are grounded in this paper and cite their sources.
Preimplantation BNP was associated with appropriate ICD therapies; leaves open its utility for risk stratification or patient selection.
OBJECTIVE: To assess prospectively whether preimplantation B-type natriuretic peptide (BNP) and C reactive protein (CRP) concentrations predict future appropriate therapies from an implantable cardioverter-defibrillator (ICD). DESIGN AND SETTING: Prospective cohort study conducted in a tertiary cardiac care centre. METHODS: 345 consecutive patients undergoing first time ICD implantation were prospectively studied. Serum BNP and CRP concentrations were obtained the day before ICD implantation. Patients were followed up with device interrogation to assess for appropriate shocks or antitachycardia pacing. Inappropriate therapies were excluded. Mean (SD) follow up was 13 (5) months. RESULTS: Patients had ischaemic (71%), primary dilated (17%), and valvar or other cardiomyopathies (12%). About half (52%) had ICDs implanted for primary prevention. Sixty three (18%) received appropriate ICD therapies. Serum creatinine, beta blocker, statin, and angiotensin converting enzyme inhibitor usage did not differ between therapy and no therapy groups. By univariate comparison, ejection fraction (p = 0.048), not taking amiodarone (p = 0.033), and BNP concentration (p = 0.0003) were risk factors for ICD therapy. However, by Cox regression multivariate analysis, only BNP above the 50th centile was a significant predictor (hazard ratio 2.19, 95% confidence interval 1.07 to 4.71, p = 0.040). Median BNP was 573 ng/l versus 243 ng/l in therapy and no therapy patients, respectively (p = 0.0003). More patients with BNP above the 50th centile (27% v 10%, p = 0.006) received ICD therapies. CONCLUSIONS: A single preimplantation BNP concentration determination is independently predictive of ICD therapies in patients with cardiomyopathies undergoing first time ICD implantation. CRP was not independently predictive of ICD therapies when compared with BNP.
No takes yet. Share an insight, caveat, or question.
Verma et al. (2005) conducted a cohort in Cardiomyopathies requiring first-time ICD implantation (n=345). Preimplantation B-type natriuretic peptide (BNP) concentration above the 50th centile vs. BNP concentration below the 50th centile was evaluated on Appropriate therapies from an implantable cardioverter-defibrillator (ICD) (HR 2.19, 95% CI 1.07 to 4.71, p=0.040). Preimplantation BNP concentration above the 50th centile independently predicted appropriate ICD therapies in patients undergoing first-time implantation.
Synapse has enriched 10 related papers. Consider them for comparative context: