Key result
Atrial pacing boosts cardiac output ~20% over ventricular pacing in recent heart transplant recipients.
Why the study?
The beneficial effects of properly sequenced atrial and ventricular activation are unclear in heart transplant recipients due to distortion of atrial morphology after transplantation.
Answers are grounded in this paper and cite their sources.
Atrial pacing improves early post-transplant hemodynamics; leaves open whether it should inform long-term pacemaker selection.
Because of the distortion of atrial morphology that occurs during cardiac allograft transplantation in humans, the beneficial effects of properly sequenced atrial and ventricular activation are unclear in these patients. To evaluate the atrial contribution to ventricular pump performance in heart transplant recipients, arterial pressure and cardiac output during pacing from either chamber were measured in nine patients 10 +/- 1 days after transplantation. Systolic, diastolic and mean systemic arterial pressures were significantly higher during atrial pacing compared with ventricular pacing: 143 +/- 23 versus 125 +/- 20 mm Hg, 73 +/- 15 versus 66 +/- 14 mm Hg and 94 +/- 17 versus 84 +/- 16 mm Hg, respectively (p less than 0.05 for all). In addition, cardiac output decreased from 5.5 +/- 1.4 to 4.6 +/- 1.5 liters/min (p less than 0.005) for atrial versus ventricular pacing. Thus, there is a significant atrial contribution to cardiac performance in patients after heart transplantation. This may have clinical implications in those patients who later require a permanent pacemaker.
No takes yet. Share an insight, caveat, or question.
Midei et al. (1990) studied Heart transplant recipients (n=9). Atrial pacing vs. Ventricular pacing was evaluated on Cardiac output (liters/min) (p=<0.005). Atrial pacing increased cardiac output compared with ventricular pacing in recent heart transplant recipients.
Synapse has enriched 10 related papers. Consider them for comparative context: