Key result
Polymer-free SES shows similar 12-month TLR rates in ACS and stable CAD.
Why the study?
The safety and efficacy of the latest generation polymer-free sirolimus-eluting stents (PF-SES) in an all-comers population with stable CAD versus ACS in France were not previously assessed.
Does the latest generation polymer-free sirolimus-eluting stent provide low rates of target lesion revascularization in patients with acute coronary syndrome compared to those with stable CAD?
Observational (n=1,456)
Yes
Does the latest generation polymer-free sirolimus-eluting stent provide low rates of target lesion revascularization in patients with acute coronary syndrome compared to those with stable CAD?
Absolute Event Rate: 2.2% vs 1.9%
p-value: p=0.7
The latest generation of polymer-free sirolimus-eluting stents demonstrate high procedural success and similarly low 12-month revascularization rates in routine clinical practice for both stable CAD and ACS.
May support polymer-free stent use in ACS; extends observational data but leaves randomized confirmation open.
Objectives. The aim of this postmarket clinical study was to assess the safety and efficacy of the latest generation polymer-free sirolimus-eluting stents (PF-SES) in an all-comers population comparing outcomes in stable coronary artery disease (CAD) versus acute coronary syndrome (ACS) in France. Background. The efficacy and safety of the first-generation PF-SES have already been demonstrated by randomized controlled trials and “all-comers” observational studies. Methods. For this all-comers observational, prospective, multicenter study, 1456 patients were recruited in 22 French centers. The primary endpoint was target lesion revascularization (TLR) rate at 12 months and secondary endpoints included major adverse cardiac events (MACE) and bleeding. Results. 895 patients had stable CAD and 561 had ACS. At 12 months, 2% of patients had a TLR, with similar rates between stable CAD and ACS (1.9% vs 2.2%, p = 0.7). The overall MACE rate was 5.2% with an expected higher rate in patients with ACS as compared to those with stable CAD (7.3% vs 3.9%, p = 0.007). The overall bleeding event rate was 4.5%, with similar rates in stable CAD as compared to ACS patients (3.8% vs 5.6%, p = 0.3). Dual antiplatelet therapy (DAPT) interruptions prior to the recommended duration occurred in 41.7% of patients with no increase in MACE rates as compared to patients who did not prematurely interrupt DAPT (3.9% vs 6.1%, p = 0.073). Conclusions. The latest generation PF-SES is associated with low clinical event rates in these all-comers patients. There was a high rate of prematurely terminated DAPT, without any effect on MACE at 12 months. This trial is registered with NCT03809715.
No takes yet. Share an insight, caveat, or question.
Landolff et al. (2023) conducted an observational in Stable coronary artery disease (CAD) or acute coronary syndrome (ACS) (n=1,456). Polymer-free sirolimus-eluting stents (PF-SES) in acute coronary syndrome vs. Polymer-free sirolimus-eluting stents (PF-SES) in stable coronary artery disease was evaluated on Accumulated target lesion revascularization (TLR) rate at 12 months (p=0.7). The latest generation polymer-free sirolimus-eluting stents were associated with an overall 12-month target lesion revascularization rate of 2.0%, which was similar between patients with acute coronary syndrome and stable coronary artery disease (2.2% vs 1.9%, p=0.7).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: