Key result
Lattice-tip ventricular ablation linked to ~0.7-V acute RV pacing threshold rise that resolves by 3 months.
Why the study?
The effects of ventricular ablation using a dual-energy lattice-tip catheter on ventricular leads of cardiac implantable electronic devices remain unclear.
Does ventricular ablation using a dual-energy lattice-tip catheter affect pacing thresholds in patients with cardiac implantable electronic devices?
Observational (n=79)
Does ventricular ablation using a dual-energy lattice-tip catheter affect pacing thresholds in patients with cardiac implantable electronic devices?
Mean Difference: 0.74 (95% CI 0.465–1.018)
Ventricular ablation using a dual-energy lattice-tip catheter can cause acute, transient pacing threshold elevations in leads near the ablation site, highlighting the need for careful peri-procedural assessment.
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“Our findings also revealed a need for caution when PFA is delivered near exposed implanted cardiac device conductors, when the catheter is maneuvered near the mitral or tricuspid valves, and when epicardial mapping may risk the trapping of fat within the catheter sphere. This early experience is encouraging, but further use and investigation will be required to optimize PFA parameters for use in ventricular substrates.”
May warrant peri-procedural threshold checks near ablation sites; leaves open long-term CIED safety with lattice-tip catheters.
Background Interactions between cardiac implantable electronic devices (CIEDs) and pulsed field ablation (PFA) have been studied mainly in the context of atrial ablation, whereas the effects of ventricular ablation using a dual‐energy lattice‐tip catheter on ventricular leads remain unclear. Methods This retrospective study involved 79 patients (mean age 60.9 ± 13.6 years) with an implantable cardioverter‐defibrillator or cardiac resynchronization therapy defibrillator who underwent ventricular ablation using a lattice‐tip catheter between February 2024 and November 2025. Pacing thresholds were assessed before, immediately after, and 3 months after ablation and standardized to a pulse width of 0.4 ms using the Lapicque–Weiss model. Marked acute threshold elevation was defined as an increase in pacing threshold of ≥ 1.0 V immediately after ablation. The anatomical relationship between ablation sites and ventricular lead locations was assessed based on the American Heart Association 17‐segment model of the left ventricle and classified as corresponding when ablation involved the same or opposing ventricular region as the lead location, adjacent when ablation involved a directly adjacent region, and remote otherwise. Results Marked acute pacing‐threshold elevation occurred in 5 of 79 right ventricular (RV) leads (6.3%) and 2 of 22 evaluable left ventricular (LV) leads (9.1%). For RV leads, marked threshold elevation occurred in 4 of 14 patients (28.6%) in the corresponding group, 1 of 19 (5.3%) in the adjacent group, and none of 46 (0%) in the remote group. When threshold change was analyzed as a continuous variable, the acute increase was significantly greater in the corresponding group (+ 0.74 V; 95%CI 0.465–1.018) than in the adjacent (+ 0.31 V; 95%CI 0.076–0.550) and remote (+ 0.007 V; 95%CI − 0.145–0.160) groups. For LV leads, the pacing threshold increased significantly after ablation involving the corresponding segment (+ 0.71 V; 95%CI 0.167–1.245), although the overall association between anatomical relationship and threshold change was not significant. At 3 months, pacing thresholds in the corresponding groups were not significantly different from baseline for either RV or LV leads. Conclusion Marked acute pacing‐threshold elevation was observed in both RV and LV leads after ventricular arrhythmia ablation using a dual‐energy lattice‐tip catheter. However, both RV and LV thresholds were no longer significantly different from baseline at 3 months. These findings suggest the importance of considering the anatomical relationship between ablation sites and CIED leads and of careful peri‐procedural assessment of lead function during ventricular ablation.
No takes yet. Share an insight, caveat, or question.
Hayasaka et al. (2026) conducted an observational in Ventricular arrhythmia in patients with cardiac implantable electronic devices (n=79). Ventricular ablation using a dual-energy lattice-tip catheter vs. Adjacent and remote ablation sites was evaluated on Acute increase in right ventricular lead pacing threshold when ablation involved the corresponding segment (MD 0.74 V, 95% CI 0.465-1.018). Ventricular ablation using a dual-energy lattice-tip catheter caused an acute increase in RV pacing thresholds at corresponding sites (MD 0.74 V; 95% CI 0.465-1.018), which resolved by 3 months.