Key result
Piggy-back device implantation appears safe with no device-device interactions in small series.
Why the study?
The usual approach to replacing cardiac devices often sacrifices existing units with good battery longevity, prompting exploration of alternative implantation methods.
Case Report (n=4)
Answers are grounded in this paper and cite their sources.
Following the publication of several landmark trials, there has been a significant increase in the cardiac device implantation. Within this population there are a number of patients who have pre-existing cardiac devices that have been placed for a number of different conditions. While the usual approach is to remove the existing unit and replace it with a new device with the removal or capping of existing lead systems, this practice often sacrifices an existing unit that still possesses good battery longevity. We explored the possibility of separating the pacing and defibrillating functions by implanting a new device on the top of the old device in a 'piggy-back fashion'. We report a series of four cases (with various indications) with differing combinations of devices. The procedure was performed safely in every one of them, and no device-device interaction was noted. Combining the new with existing units in a 'piggy-back' manner may be a safe and cost-effective technique in the selected cases.
No takes yet. Share an insight, caveat, or question.
Kanjwal et al. (2007) conducted a case report in Pre-existing cardiac devices needing new device implantation (n=4). Piggy-back pacing (implantation of a new device on top of an old device) was evaluated on Safety and device-device interaction. Implanting a new pacemaker or defibrillator on top of an existing device in a piggy-back fashion was performed safely in 4 cases with no device-device interactions noted.
Synapse has enriched 10 related papers. Consider them for comparative context: