Key result
Electrocardiograph devices yield significantly different ECG durations and amplitudes despite normal-range values in sheep.
Why the study?
In sheep, electrocardiographic values vary by breed, age, and capture method, but values obtained across different computerized devices had not been compared.
Do two different computerized electrocardiography devices (Incardio Agile vs. DL660) produce different electrocardiographic variable values in healthy adult female Suffolk sheep?
Do two different computerized electrocardiography devices (Incardio Agile vs. DL660) produce different electrocardiographic variable values in healthy adult female Suffolk sheep?
Different computerized electrocardiograph models yield statistically different wave duration and amplitude measurements in adult Suffolk sheep due to variations in tracing capture sensitivity, though values largely remain within established normal ranges for the species.
Answers are grounded in this paper and cite their sources.
Device differences in Suffolk sheep ECGs require cautious interpretation; leaves open device-specific reference standardization.
Computerized electrocardiography is a method widely used in veterinary medicine in several species and currently, with the availability of several devices in the market. In sheep, studies have shown variations in electrocardiographic values according to race, age, and methods of capture, but the values obtained in different computerized devices have not been compared. Thus, the present study sought to determine and compare the values of electrocardiographic variables between two computerized devices in healthy sheep. Fifty sheep (n = 50), Suffolk breed, female, launched, adult, non-pregnant, with average age and weight of two years and 50kg, respectively, were studied properly selected and clinically fit. The electrocardiographic tracings were obtained in two computerized electrocardiography devices (Apparatus A and B) individually and subsequently, with subsequent recording and interpretation of the tracings. The results showed a predominance of sinus rhythm in 78% (n = 39) and 22% (n = 11) with sinus arrhythmia in both devices. Significant differences among the devices (p <0.05) were identified in the duration variables (ms), such as P wave, P-R interval and segment, QRS complex, Q-T interval, S-T segment and T wave; and in the P-wave and R-wave amplitudes (mV). However, the values obtained were within the normal ranges recommended for the species, regardless of the capture method used. Thus, it is possible to conclude the presence of different values among the computerized electrocardiographs, suggesting variations in the tracking uptake sensitivity among the devices in released Suffolk sheep.
No takes yet. Share an insight, caveat, or question.
Franco et al. (2020) studied Healthy adult Suffolk sheep undergoing electrocardiographic evaluation (n=50). Computerized Electrocardiograph Device A (ICAV1.2 - Incardio Agile - 6 leads) vs. Computerized Electrocardiograph Device B (DL660 - 12 leads) was evaluated on Electrocardiographic duration (ms) and amplitude (mV) variables in lead D2 (p=<0.05). Computerized electrocardiograph devices A and B produced different duration and amplitude values in healthy Suffolk sheep, though remaining within species normal ranges.