Key result
High implantation depth-to-membranous septum ratio linked to ~6-fold higher post-TAVR conduction disturbances.
Why the study?
Conduction disturbances are a common complication after TAVR with self-expanding valves, and the value of the implantation depth-to-membranous septum ratio (DIMS) and a corresponding risk model for predicting new-onset conduction disturbances with repositionable self-expanding valves required evaluation.
Does the implantation depth–to–membranous septum ratio (DIMS) predict new-onset conduction disturbances within 30 days in patients with severe symptomatic aortic stenosis undergoing TAVR with repositionable self-expanding valves?
Observational (n=74)
No
Does the implantation depth–to–membranous septum ratio (DIMS) predict new-onset conduction disturbances within 30 days in patients with severe symptomatic aortic stenosis undergoing TAVR with repositionable self-expanding valves?
In patients undergoing TAVR with repositionable self-expanding valves, an implantation depth–to–membranous septum ratio (DIMS) >41.33%, combined with baseline QRS ≥120 ms and aortic angulation ≥50°, independently predicts 30-day conduction disturbances with moderate discriminative ability (AUC 0.78).
Answers are grounded in this paper and cite their sources.
DIMS ratio may support conduction risk stratification in self-expanding TAVR; leaves open whether it improves outcomes in prospective studies.
Objective: Conduction disturbances remain a common complication following transcatheter aortic valve replacement (TAVR), particularly with self-expanding valve systems. This study aimed to evaluate the association between the implantation depth–to–membranous septum ratio (DIMS) and the occurrence of new-onset conduction disturbances, and to develop a hypothesis-generating risk model in patients undergoing TAVR with repositionable self-expanding valves. Methods: In this prospective observational study, 74 patients with severe symptomatic aortic stenosis undergoing TAVR were included. The primary endpoint was the occurrence of new-onset conduction disturbances within 30 days, defined as a composite of atrioventricular block, new left bundle branch block (LBBB), or PR interval prolongation. Multivariable logistic regression analysis was performed to identify independent predictors. A predictive model was constructed using significant variables and internally validated using bootstrap resampling techniques. Results: Conduction disturbances occurred in 25.6% of patients, with 17.5% classified as persistent. The DIMS ratio was independently associated with post-TAVR conduction disturbances. When combined with baseline QRS duration and aortic angulation, the predictive model demonstrated moderate discriminative ability (AUC 0.78). A hypothesis-generating risk score incorporating these variables showed high specificity but limited sensitivity. Predictive performance should be interpreted cautiously due to the limited number of outcome events and absence of external validation. Conclusion: The DIMS ratio is associated with conduction disturbances following TAVR using repositionable self-expanding valves. Integration of anatomical and electrocardiographic parameters may enhance risk stratification; however, these findings are exploratory and require validation in larger, multicenter cohorts before clinical application.
No takes yet. Share an insight, caveat, or question.
Abouelnour et al. (2026) conducted an observational in Severe symptomatic aortic stenosis undergoing TAVR (n=74). Implantation depth-to-membranous septum ratio (DIMS) >41.33% vs. DIMS ≤41.33% was evaluated on New-onset conduction disturbances within 30 days (composite of atrioventricular block, new left bundle branch block, or PR interval prolongation) (OR 6.0, 95% CI 1.2-30.3, p=0.029). An implantation depth-to-membranous septum ratio >41.33% independently predicted 30-day new-onset conduction disturbances after TAVR.