Key result
Nios-V soft-processor architecture enables real-time ECG denoising with reduced power and hardware consumption.
Why the study?
Embedded architectures introduced to filter ECG signals still face challenges in processing frequencies, hardware consumption, power, and time due to the elevated computational complexity of the algorithms.
Does a Nios-V soft-processor based system-on-chip architecture implementing the DWT-ADTF process improve real-time ECG signal denoising and hardware efficiency on MIT-BIH Arrhythmia database recordings?
Does a Nios-V soft-processor based system-on-chip architecture implementing the DWT-ADTF process improve real-time ECG signal denoising and hardware efficiency on MIT-BIH Arrhythmia database recordings?
A Nios-V soft-processor FPGA architecture enables real-time, sample-by-sample DWT-ADTF ECG denoising at 50 MHz in 0.41 ms without permanent data storage, supporting integration into low-power ambulatory ECG monitoring systems.
Answers are grounded in this paper and cite their sources.
May enable portable real-time ECG monitoring; leaves open prospective clinical validation.
Several are the embedded architectures that have been introduced to filter ECG signals. However, processing frequencies, consumed hardware, power, and time are still challenging. The elevated computational complexity of the considered algorithms is the major reason. The DWT-ADTF is a recent low-computational complexity process that has been introduced for ECG denoising. This paper proposes an FPGA-based software architecture providing real-time ECG signal filtering with low frequency and power consumption. Our system consists of a Nios-V soft processor interconnected to 69.34 KB of RAM through the AXI4 and Avalon interfaces. The Nios-V is used thanks to its hardware flexibility and five-stage pipeline. Using the C language, our architecture is designed to take advantage of this hardware capacity. The data is loaded and treated through a sample-by-sample and shifting approach. Consequently, the processor’s pipeline is well-derived for real-time processing, and no permanent data storage is needed. Thus, under only 50 MHz, our architecture ensures real-time data loading and processing while consuming less hardware and power. The denoising quality is assessed using the average MSE, PRD, and SNRimp. Several noise types and ECG signals from the MIT-BIH Arrhythmia database are involved. The processing frequency and the consumed hardware, power, and time evaluate the efficiency and performance of our architecture. Promising results are achieved compared to other works, which makes our architecture appropriate for real-time ECG monitoring systems. Further, its lower consumption favors more ECG processing integration.
No takes yet. Share an insight, caveat, or question.
Bendahane et al. (2026) studied ECG signal noise contamination (n=8). Nios-V soft-processor based system-on-chip architecture implementing the DWT-ADTF algorithm vs. Original MATLAB DWT-ADTF software implementation and related ECG filtering architectures was evaluated on ECG denoising quality (average MSE, PRD, and SNRimp) and architectural performance (execution time, hardware consumption, and processing frequency). The Nios-V soft-processor DWT-ADTF architecture achieved real-time ECG denoising in 0.41 ms at 50 MHz using 69.34 KB RAM, 3,945 registers, 9% logic elements, and 6% DSPs.
Synapse has enriched 10 related papers. Consider them for comparative context: