An absorbance detector for high-performance liquid chromatography (HPLC) wasdeveloped using a commercially available light emitting diode (LED) at 254 nm. Use of theLED was investigated due to its low output power fluctuation, which should minimize noiseand maximize performance. This detector has been characterized and used to performseveral separations of relevance to the atmospheric chemistry of airborne particulatematter, specifically, brown carbon (BrC) or humic-like substances (HULIS). The study ofatmospheric BrC can be aided by sensitive and general detection schemes, such asabsorbance detection. Owing to the exceptional output stability of the LED, the absorbancedetector exhibited 3σ detection limits of 130 nmol/L (18 ppb) for the chromatography of4-hydroxybenzoic acid and the detector itself demonstrated noise-equivalent absorptionof approx. 180 µAU. The detector was applied to the separation of products resulting fromthe photo-Fenton reaction of guaiacol. No fewer than 15 individual peaks are noted in theresulting chromatogram. The sensitivity provided is a valuable tool for the chemical analysisof complex samples requiring chromatography.