Why the study?
Previous methods for in vivo determination of left ventricular mass required ionizing radiation, complicated geometric assumptions, or invasive techniques, prompting evaluation of gated nuclear magnetic resonance imaging as an alternative.
Population
Anesthetized animals including 9 dogs and 2 cats
Comparison
In vivo gated nuclear magnetic resonance imaging calculated LV mass vs actual postmortem LV mass
Design
Animal validation study
Answers are grounded in this paper and cite their sources.
Demonstrates accurate in vivo measurement of left ventricular mass in animals using gated nuclear magnetic resonance imaging, highlighting a noninvasive alternative to ionizing radiation.
Alterations of left ventricular mass occur in a variety of congenital and acquired heart diseases. In vivo determination of left ventricular mass, using several different techniques, has been previously reported. Problems inherent in some previous methods include the use of ionizing radiation, complicated geometric assumptions and invasive techniques. We tested the ability of gated nuclear magnetic resonance imaging to determine in vivo left ventricular mass in animals. By studying both dogs (n = 9) and cats (n = 2) of various sizes, a broad range of left ventricular mass (7 to 133 g) was examined. With a 0.5 tesla superconducting nuclear magnetic resonance imaging system the left ventricle was imaged in the transaxial plane and multiple adjacent 10 mm thick slices were obtained. Endocardial and epicardial edges were manually traced in each computer-displayed image. The wall area of each image was determined by computer and the areas were summed and multiplied by the slice thickness and the specific gravity of muscle, providing calculated left ventricular mass. Calculated left ventricular mass was compared with actual postmortem left ventricular mass using linear regression analysis. An excellent relation between calculated and actual mass was found (r = 0.95; SEE = 13.1 g; regression equation: magnetic resonance mass = 0.95 X actual mass + 14.8 g). Intraobserver and interobserver reproducibility were also excellent (r = 0.99). Thus, gated nuclear magnetic resonance imaging can accurately determine in vivo left ventricular mass in anesthetized animals.
No takes yet. Share an insight, caveat, or question.
Florentine et al. (1986) studied this question.