This study introduces a successfully manufactured U-shaped optical fiber sensor integrated with a microfluidics system to detect varying glucose concentrations. The outlines the fabrication process of the microfluidics system, accomplished using a CO2 laser on polymethyl methacrylate (PMMA) material. Through three cycles of experiments involving different glucose concentrations, the results indicate a consistent trend in both resonance wavelength and transmission loss, showing an increase as glucose concentration rises from 5% to 50%. The system demonstrates a high wavelength sensitivity coefficient of 0.114 nm/% with R 2 = 0.95 and a notable transmission loss sensitivity of 0.09 dB/% and R 2 = 0.94. This advanced system proves to be effective in achieving rapid, sensitive, and cost-efficient glucose monitoring. Consequently, the microfluidics system is deemed suitable for its intended purpose.
No takes yet. Share an insight, caveat, or question.
Hsu et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: