Key result
Upsizing artificial aortic valves to 25 mm improves hemodynamics, while 27 mm raises thrombosis risk.
Why the study?
Does upsizing Medtronic Avalus heart valves improve haemodynamics and reduce thrombosis risk in computational fluid dynamics models?
Does upsizing Medtronic Avalus heart valves improve haemodynamics and reduce thrombosis risk in computational fluid dynamics models?
Computational fluid dynamics modeling suggests that moderate upsizing to a 25 mm Medtronic Avalus valve improves hemodynamics, while further upsizing to 27 mm increases shear stress, turbulence, and thrombosis risk.
OBJECTIVES: Prosthesis-patient mismatch (PPM) and tilted implantation in aortic valve replacement can significantly affect haemodynamics, increasing wall shear stress (WSS) and thrombosis risk. This study uses computational fluid dynamics (CFD) to assess the impact of valve size and implantation angle on blood flow in Medtronic Avalus heart valves. METHODS: Computational fluid dynamics (CFD) simulations were conducted using Ansys Fluent to evaluate 3 Medtronic Avalus valve sizes (23 mm, 25 mm, 27 mm) under 2 leaflet-opening configurations. The simulations incorporated physiological boundary conditions, including a steady inlet velocity of 0.3 m/s and a no-slip wall condition. Wall shear stress (WSS) and general shear stress (SS) thresholds were analysed to assess thrombosis risk. RESULTS: The 25 mm valve reduced the WSS-affected area by 20.83% and the shear stress-affected area by 7.64% compared to the 23 mm valve, demonstrating improved haemodynamics. In contrast, the 27 mm valve increased the WSS-affected area by 41.25% and the shear stress-affected area by 11.26%, with elevated turbulence and recirculation zones near the aortic wall, increasing thrombosis risk. CONCLUSIONS: Moderate upsizing to a 25 mm valve improves haemodynamic performance without introducing adverse flow effects, while further upsizing to 27 mm increases stress and turbulence, raising thrombosis risk.
No takes yet. Share an insight, caveat, or question.
Berger et al. (2026) studied Aortic valve replacement. 25 mm and 27 mm Medtronic Avalus valves vs. 23 mm Medtronic Avalus valve was evaluated on Wall shear stress (WSS) and general shear stress (SS) affected areas.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: