Key result
Tilting of bileaflet valves by 20 degrees resulted in a significant size-dependent increase in mean pressure gradient and impaired overall hemodynamics.
Why the study?
Does a tilted implantation position impair the hemodynamic performance of standard bileaflet mechanical valves in an in vitro mock circulatory system?
Does a tilted implantation position impair the hemodynamic performance of standard bileaflet mechanical valves in an in vitro mock circulatory system?
Tilting of bileaflet mechanical valves significantly impairs systolic and diastolic hemodynamics, particularly at a 90-degree rotation, suggesting such positions should be avoided or surgically corrected.
Answers are grounded in this paper and cite their sources.
May warrant caution with tilted bileaflet valve positions; hypothesis-generating from in vitro data and leaves open clinical relevance.
OBJECTIVES: Severe sclerosis of the native aortic annulus can result in a tilted implantation position of mechanical prostheses. In this study, the effects of tilting and rotation on the hemodynamic performance of standard bileaflet valves were assessed in an extracorporeal mock circulatory system. METHODS: A pulsatile mock circulation driven by a Berlin Heart system was developed. Main physiological components of the human circulation were mimicked. SJM-AHPJ prostheses (21, 23, 25 mm) were mounted in an artificial aortic root containing physiologically oriented coronary ostia. All experiments were performed under constant conditions (stroke volume 60 ml, heart rate 70 bpm, systolic pressure 130 mmHg). Hydrostatic pressures were measured via fluid-filled catheters, transvalvular flow by ultrasonic probes. Data were digitally recorded at 50 Hz. Multiple pressure, volume, energy, and dimension parameters were derived off-line. Each valve was tested in a 0 degrees (untilted) versus 20 degrees (tilted) position at three axial rotation angles (0 degrees, 45 degrees, 90 degrees ). Tilting was performed independent of rotation by elevation of the prosthesis in the non-coronary sinus. RESULTS: In all valves and all rotation angles, tilting resulted in a size-dependent significant increase of mean pressure gradient (range, 28-35% [21 mm valve], 59-96% [23 mm valve], 124-220% [25 mm valve]), valvular resistance (39-51, 84-121, 177-332%), regurgitation volume (84-148, 32-131, 93-118%), and systolic energy loss (113-146, 30-132, 69-213%), as well as a decrease of total stroke volume (2-5, 0-11, 3-10%), effective stroke volume (6-11, 9-14, 14-22%), cardiac output (6-11, 8-14, 13-22%), and effective opening area (16-24, 32-37, 47-57%). The strongest impairment of hemodynamic performance was seen at 90 degrees rotation with reference to total and effective stroke volume, cardiac output, mean pressure gradient, and regurgitation fraction. CONCLUSIONS: Tilting of bileaflet valves resulted in a significant impairment of systolic and diastolic hemodynamics. Superiority of larger valves diminished in the tilted position. The strongest tilting effect was seen at 90 degrees rotation. Such a position should therefore be avoided or surgically corrected by rotating the valve.
No takes yet. Share an insight, caveat, or question.
Martin Hartrumpf (2003) studied Aortic valve replacement hemodynamics. Tilted implantation position (20 degrees) vs. Untilted position (0 degrees) was evaluated on Hemodynamic performance including mean pressure gradient, valvular resistance, regurgitation volume, and stroke volume. Tilting of bileaflet valves by 20 degrees resulted in a significant size-dependent increase in mean pressure gradient and impaired overall hemodynamics.