Why the study?
Does the subcutaneous implantable cardioverter-defibrillator (S-ICD) effectively convert ventricular tachyarrhythmia in patients requiring primary or secondary prevention of sudden cardiac death?
Does the subcutaneous implantable cardioverter-defibrillator (S-ICD) effectively convert ventricular tachyarrhythmia in patients requiring primary or secondary prevention of sudden cardiac death?
In a clinical setting, the S-ICD demonstrated high overall shock efficacy but low first-shock efficacy, indicating that ineffective shock delivery may occur despite successful intraoperative testing.
Answers are grounded in this paper and cite their sources.
May support multiple-shock protocols for S-ICD in practice; leaves open impact on outcomes in prevention cohorts.
BACKGROUND: Recently, subcutaneous implantable cardioverter-defibrillator (S-ICD) has become available. The aim of our study was to assess the efficacy of S-ICD in a clinical setting. METHODS AND RESULTS: Between June 2010 and July 2011, 40 consecutive patients (42 ± 15 years; body mass index, 27 ± 6 kg/m(2); left ventricular ejection fraction, 47 ± 15%; 28 men) received an S-ICD for primary (n=17) or secondary prevention (n=23 [58%]) at 3 institutions in Germany. Intraoperative defibrillation efficacy testing failed in 1 patient with severely reduced left ventricular ejection fraction; testing was effective in all other patients. All episodes stored in the S-ICD were analyzed for appropriate and inappropriate detection, as well as effective shock delivery to convert ventricular tachyarrhythmia into sinus rhythm. During a median follow-up of 229 (interquartile range, 116-305) days, 4 patients experienced 21 episodes, with correct detection of ventricular tachyarrhythmia and subsequent shock therapy. A total of 28 shocks were delivered in these 4 patients. Mixed logistic regression modeling revealed a shock efficacy of 96.4% (95% CI, 12.8%-100%). The efficacy of first shocks, however, was only 57.9% (95% CI, 35.6%-77.4%). Four episodes were incorrectly classified as ventricular tachyarrhythmia, which led to inappropriate shock delivery in 2 patients. CONCLUSIONS: Ineffective shock delivery may occur in patients with S-ICD, even after successful intraoperative testing. Multicenter trials are required with close monitoring of safety and efficacy end points to identify patients who may be at risk for shock failure.
No takes yet. Share an insight, caveat, or question.
Aydın et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: