Key result
Bidirectional Evolution R/L sheath links to ~26% higher complete RV lead extraction success vs first-generation sheaths.
Why the study?
Mechanically powered lead extraction is effective, but the safety of dissecting adhesions up to the right ventricle lead tip at the myocardium was unclear.
Does the bidirectional Evolution R/L sheath improve complete success rates and safety compared to the first-generation Evolution sheath in patients undergoing RV-lead extraction?
Observational (n=185)
No
Does the bidirectional Evolution R/L sheath improve complete success rates and safety compared to the first-generation Evolution sheath in patients undergoing RV-lead extraction?
The bidirectional Evolution R/L sheath is safe and highly effective for extracting right ventricular leads, even when adhesions extend up to the myocardium.
Answers are grounded in this paper and cite their sources.
May enhance RV lead extraction success; hypothesis-generating and should not yet change practice.
AIMS: Goal of Transvenous Lead Extraction (TLE) is complete removal of all targeted leads, without complications. Despite counter traction manoeuvres, efficacy rates are often hampered by broken right ventricle lead (RV-lead) tips. Mechanically powered lead extraction (Evolution sheath) is effective, however safety of dissection up to the lead tip is unclear. Therefore, we examined the feasibility and safety of RV-lead extraction requiring dissection up to the myocardium. METHODS AND RESULTS: From 2009 to 2018, all TLE in the Isala Heart Centre (Zwolle, The Netherlands) requiring the hand-powered mechanical Evolution system to extract RV-leads (n = 185) were examined from a prospective registry. We assessed 4 groups: TLE with the first generation Evolution (n = 43) with (A1,n = 18) and without (A2,n = 25) adhesions up to the myocardium and TLE with the Novel R/L type (n = 142) of sheath with (B1, n = 59) and without (B2, n = 83) adhesions up to the myocardium. Complete success rate in Group B was significantly higher than group A (96.5 vs 76.7%, p = 0.0354). When comparing the patients with adhesions up to the myocardium, total complete success is higher in the R/L group (61.1% vs 90.5%, p = 0.0067). There were no deaths. Overall major complication rates were low (2/185; 1.1%) and there was no statistically significant difference in major and minor complications between the two groups. CONCLUSION: Extraction strategy with the bidirectional Evolution R/L sheath for right ventricular leads with adhesions up to the myocardium is safe and feasible.
No takes yet. Share an insight, caveat, or question.
Witte et al. (2021) conducted an observational in Transvenous Lead Extraction (TLE) of right ventricular leads (n=185). Novel R/L type Evolution sheath vs. First generation Evolution sheath was evaluated on Complete success rate (p=0.0354). Extraction of right ventricular leads using the bidirectional Evolution R/L sheath achieved a higher complete success rate than the first-generation sheath.
Synapse has enriched 10 related papers. Consider them for comparative context: