Key result
Active fixation RV leads with retracted helices linked to ~59% lower odds of difficult extraction.
Why the study?
The impact of lead fixation mechanism on extractability is poorly characterized.
Does right ventricular lead fixation mechanism and helix retraction status affect the technical difficulty of transvenous lead extraction?
Cohort (n=408)
No
Does right ventricular lead fixation mechanism and helix retraction status affect the technical difficulty of transvenous lead extraction?
Successful helical retraction is a critical determinant of lead extraction difficulty, as unretracted active fixation leads pose technical challenges comparable to historically difficult passive fixation leads.
Answers are grounded in this paper and cite their sources.
May ease TLE planning when retraction succeeds; supports helix status assessment but leaves practice change open.
Background: The impact of lead fixation mechanism on extractability is poorly characterized. Objective: We aimed to compare the technical difficulty of transvenous lead extraction (TLE) of active vs passive fixation right ventricular (RV) leads. Methods: A total of 408 patients who underwent RV TLE by a single expert electrophysiologist at Oregon Health & Science University between October 2011 and June 2022 were identified and retrospectively analyzed; 331 (81%) had active fixation RV leads and 77 (19%) had passive fixation RV leads. The active fixation cohort was further stratified into those with successfully retracted helices (n = 181) and failed helix retraction (n = 109). A numerical system (0-9) devised using 6 procedural criteria quantified a technical extraction score (TES) for each RV TLE. The TES was compared between groups. The TLE success rate of the entire cohort was >97%, with a major complication rate of 0.5%. Conclusion: TLE of active fixation leads where helical retraction is achieved presents fewer technical challenges than does passive fixation RV lead extraction; however, if the helix cannot be retracted, active and passive TLE procedures present similar technical challenges.
No takes yet. Share an insight, caveat, or question.
Frazer et al. (2023) conducted a cohort in Right ventricular lead requiring transvenous extraction (n=408). Active fixation RV lead with successfully retracted helix vs. Passive fixation RV lead was evaluated on High technical extraction difficulty (Technical Extraction Score > 4) (Adjusted OR 0.41, 95% CI 0.19-0.87, p=0.008). Active fixation right ventricular leads with successfully retracted helices had 59% lower adjusted odds of a highly difficult extraction procedure compared to passive fixation leads.
Synapse has enriched 10 related papers. Consider them for comparative context: