Key result
Active fixation of a permanent pacing lead inside the right inferior ganglionated plexus using a novel guiding catheter allowed lead delivery within 40 minutes in dogs with atrial fibrillation.
Why the study?
Does implantation of a permanent pacing lead inside the right inferior ganglionated plexus reduce ventricular rate in dogs with atrial fibrillation?
Does implantation of a permanent pacing lead inside the right inferior ganglionated plexus reduce ventricular rate in dogs with atrial fibrillation?
A novel transvenous approach for implanting a permanent pacing lead in the right inferior ganglionated plexus is feasible and allows for ventricular rate control during atrial fibrillation in a canine model.
Answers are grounded in this paper and cite their sources.
Does not change AF practice; hypothesis-generating for ganglionated plexus pacing to control ventricular rate in humans.
AIMS: Ventricular rate control is considered as an initial choice of therapy in many patients with atrial fibrillation (AF). We could previously show that electrostimulation of the right inferior ganglionated plexus (RIGP), which supplies the AV node, instantly decreases ventricular rate during AF. This study describes the development of a technique to reliably implant a chronic lead inside the RIGP. METHODS AND RESULTS: In nine mongrel dogs with AF, the RIGP was identified by neuromapping with probatory high-frequency stimulation (20 Hz) over steerable electrode catheters until a significant ventricular rate slowing was achieved. Then an active fixation, permanent pacemaker lead was fixed closed to the mapping catheter left in place as anatomical marker. Initially (n = 4) available guiding catheters and steerable lead stylets were employed to navigate and anchor the lead, which resulted in repetitive screw-in attempts. Therefore, a guiding catheter was developed, which allowed angiography, lead advancement through its lumen, and probatory neurostimulation over its tip. This tool allowed lead delivery within 40 min (n = 5). Neurostimulation via the permanent lead elicited negative dromotropic effects with stimulation frequency, voltage, and impulse duration as determinants of stimulation efficacy. CONCLUSION: Active fixation of a permanent pacing lead inside the RIGP is feasible without thoracotomy. Thereby, ventricular rate control during AF can be achieved with stimulus voltages applied for myocardial electrostimulation.
No takes yet. Share an insight, caveat, or question.
Gemein et al. (2009) studied Atrial fibrillation (n=9). Active fixation of a permanent pacing lead inside the right inferior ganglionated plexus (RIGP) was evaluated on Feasibility of lead implantation and ventricular rate slowing. Active fixation of a permanent pacing lead inside the right inferior ganglionated plexus using a novel guiding catheter allowed lead delivery within 40 minutes in dogs with atrial fibrillation.