Does non-invasive oscillometric blood pressure measurement using the Omron HEM-703CP monitor agree with direct radial artery measurements?
Does non-invasive oscillometric blood pressure measurement using the Omron HEM-703CP monitor agree with direct radial artery measurements?
The Omron HEM-703CP monitor provides non-invasive blood pressure and pulse measurements that correlate excellently with direct radial artery measurements.
Answers are grounded in this paper and cite their sources.
Supports limited agreement of oscillometric BP with invasive measures; leaves open larger validation before clinical substitution.
Non-invasive oscillometric blood pressure and pulse measured by an Omron HEM-703CP monitor were compared with arterial values obtained from direct measurements of the radial artery. An excellent correlation and agreement was found between the two methods (systolic r = 0.99; diastolic r = 0.97; pulse r = 0.99), although there was some variability among individual subjects. The range of difference between them was 0 to 10 mmHg for systolic and -6 to +5 mmHg for diastolic pressures. When tested on the bench using the Metron QA-1280 non-invasive blood pressure analyser the HEM-703CP monitor rarely exhibited errors exceeding 2-3 mmHg over a measurement range of 50-200 mmHg.