Why the study?
Is a new bidirectional transvenous defibrillation endocardial lead system safe and effective in patients with ventricular tachycardia or fibrillation?
Is a new bidirectional transvenous defibrillation endocardial lead system safe and effective in patients with ventricular tachycardia or fibrillation?
The new bidirectional transvenous defibrillation endocardial lead system demonstrated acceptable defibrillation thresholds and safety in the majority of patients, though 8 of 47 required epicardial leads.
Answers are grounded in this paper and cite their sources.
Should not yet alter lead selection in ICD candidates; hypothesis-generating for bidirectional transvenous systems in VT/VF.
The clinical efficacy and safety of a new bidirectional transvenous defibrillation endocardial lead system (ELS) was studied in 39 patients with ventricular tachycardia (VT) or fibrillation (VF). There were 28 patients with coronary disease and 11 patients with nonischemic VT/VF. Fourteen patients received the ELS combined with antitachycardia pacing devices (Ventak PRx 1700, CPI) and 25 patients with the Ventak P or P2 (CPI). Implantation of the ELS was attempted in 47 patients. Intraoperatively, the mean defibrillation threshold (DFT) was > 25 joules in five patients and no reliable ELS position was found in three other patients. These eight patients underwent thoracotomy and epicardial leads implantation. The mean DFT was < or = 20 joules in all 39 patients and the mean DFT was 18 joules. During the mean follow-up of 8 +/- 2 months two patients (5%) died suddenly. Complications occurred in two patients (5%).
No takes yet. Share an insight, caveat, or question.
Trappe et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: