Why the study?
Does a 50% tilt in biphasic defibrillation shocks reduce the defibrillation threshold compared to other tilts in patients undergoing transvenous defibrillator implant?
Population
33 patients undergoing transvenous defibrillator implant
Comparison
Biphasic defibrillation shocks with 50% tilt vs Biphasic defibrillation shocks with 40%, 65%…
Design
RCT, Tested in random fashion/random order
Authors
Loading...
Supports 50% tilt biphasic shocks to minimize defibrillation energy at ICD implant; extends randomized waveform optimization data.
Does a 50% tilt in biphasic defibrillation shocks reduce the defibrillation threshold compared to other tilts in patients undergoing transvenous defibrillator implant?
Biphasic shocks with a 50% tilt require significantly less energy for defibrillation compared to 65% and 80% tilts during transvenous defibrillator implantation.
Natale et al. (1996) studied this question.