Key result
A tuned defibrillation waveform significantly lowered the delivered energy defibrillation threshold compared to a 50/50% tilt waveform (7.3 vs 8.7 J; P=0.01).
Why the study?
Does a tuned defibrillation waveform reduce defibrillation thresholds compared to a 50/50% tilt waveform in patients with implantable cardioverter defibrillators?
RCT (n=34)
random order
Yes
Does a tuned defibrillation waveform reduce defibrillation thresholds compared to a 50/50% tilt waveform in patients with implantable cardioverter defibrillators?
Absolute Event Rate: 7.3% vs 8.7%
p-value: p=0.01
A tuned defibrillation waveform significantly lowers energy and voltage defibrillation thresholds compared to a standard 50/50% tilt biphasic waveform.
Supports tuned waveforms for lower-energy ICD programming; extends biphasic optimization evidence from RCTs.
INTRODUCTION: A superior performance of a tuned waveform based on duration using an assumed cardiac membrane time constant of 3.5 ms and of a 50/50% tilt waveform over a standard 65/65% tilt waveform has been documented before. However, there has been no direct comparison of the tuned versus the 50/50% tilt waveforms. METHODS: In 34 patients, defibrillation thresholds (DFTs) for tuned versus 50/50% tilt waveforms in a random order were measured by using the optimized binary search method. High voltage lead impedance was measured and used to select the pulse widths for tuned and 50/50% tilt defibrillation waveforms. RESULTS: Delivered energy (7.3 +/- 4.6 J vs 8.7 +/- 5.3 J, P = 0.01), stored energy (8.2 +/- 5.1 J vs 9.7 +/- 5.6 J, P = 0.01), and delivered voltage (405.9 +/- 121.7 V vs 445.0 +/- 122.6 V, P = 0.008) were significantly lower for the tuned than for the 50/50% tilt waveform. In four patients with DFT >/= 15 J, the tuned waveform lowered the mean energy DFT by 2.8 J and mean voltage DFT by 45 V. For all patients, the mean peak delivered energy DFT was reduced from 29 J to 22 J (24% decrease). Multiple regression analysis showed that a left ventricular ejection fraction < 20% is a significant predictor of this advantage. CONCLUSION: Energy and voltage DFTs are lowered with an implantable cardioverter defibrillator that uses a tuned waveform compared to a standard 50% tilt biphasic waveform.
No takes yet. Share an insight, caveat, or question.
NATARAJAN et al. (2007) conducted an RCT in Defibrillation threshold testing (n=34). Tuned defibrillation waveform vs. 50/50% tilt defibrillation waveform was evaluated on Delivered energy defibrillation threshold (J) (p=0.01). A tuned defibrillation waveform significantly lowered the delivered energy defibrillation threshold compared to a 50/50% tilt waveform (7.3 vs 8.7 J; P=0.01).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: