Why the study?
What are the optimal strength-duration parameters for ventricular defibrillation in dogs using capacitor-discharge and half-sinusoidal pulses?
What are the optimal strength-duration parameters for ventricular defibrillation in dogs using capacitor-discharge and half-sinusoidal pulses?
Half-sinusoidal current waveforms require less energy for ventricular defibrillation in dogs, with an optimal duration of 1.25 to 8.3 msec for minimum energy.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for energy-efficient defibrillation waveforms in animals; leaves open translation to human clinical practice.
The threshold voltage and current necessary for ventricular defibrillation was measured in dogs using capacitor-discharge and half-sinusoidal pulses of current of various durations. From these quantities the electrical strength required for defibrillation was determined and expressed as energy, peak current, and charge per gram of heart. The optimum current duration depends on the criterion selected. It was found that with half-sinusoidal current, less energy, current, and charge were required; this type of waveform approximates that delivered by capacitor-inductor defibrillators. For minimum energy with this waveform, the optimum duration range is 1.25 to 8.3 msec and the energy required is approximately 7 mw-sec per gram of heart. For minimum current, the optimum duration is 8.3 msec; with this duration the peak half-sinusoidal current is about 30 mamp per gram of heart. There is no clearly defined duration for minimum charge for capacitor-discharge or half-sinusoidal current.
No takes yet. Share an insight, caveat, or question.
Geddes et al. (1970) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: