Why the study?
Does the duration of rectangular electrical waveforms affect the safety margin for functional depression during defibrillation in isolated canine hearts?
Population
8 isolated, perfused, contracting canine hearts
Comparison
Defibrillation with rectangular electrical… vs Comparison among the 4 different waveform…
Design
Preclinical
Follow-up
immediately after the shocks
Authors
Loading...
Shorter 2-msec waveforms reduce safety margin for cardiac depression in canine hearts; leaves open optimal duration for clinical defibrillation.
Does the duration of rectangular electrical waveforms affect the safety margin for functional depression during defibrillation in isolated canine hearts?
In an isolated canine heart model, a 2-msec rectangular defibrillation waveform had a significantly lower margin of safety for functional cardiac depression compared to 5-, 10-, and 20-msec waveforms.
Niebauer et al. (1983) studied this question.