Key result
Non-invasive haemodynamic measurement using finger arterial pressure waveform analysis showed adequate concordance with invasive Swan-Ganz catheterization (variability coefficients 11-18%).
Why the study?
Does non-invasive finger arterial pressure waveform analysis provide concordant hemodynamic measurements compared to invasive Swan-Ganz catheterization in patients with advanced heart failure?
Observational (n=25)
Does non-invasive finger arterial pressure waveform analysis provide concordant hemodynamic measurements compared to invasive Swan-Ganz catheterization in patients with advanced heart failure?
Answers are grounded in this paper and cite their sources.
Non-invasive hemodynamic monitoring using finger arterial pressure waveform analysis shows adequate concordance with invasive Swan-Ganz catheterization in patients with advanced heart failure.
AIM: Measurement of haemodynamic parameters using a Swan-Ganz catheter is of clinical importance in patients with advanced heart failure; however, its applicability is limited due to its invasiveness. The aim of the study was to estimate the concordance between invasive and non-invasive measurements of haemodynamic parameters in patients with advanced heart failure. METHODS AND RESULTS: We examined 25 patients with advanced heart failure (20 men, age: 64 ± 11 years, New York Heart Association class III/IV: 88/12%, left ventricular ejection fraction: 37 ± 20%), 13 (52%) demonstrated decompensated heart failure. Resting haemodynamic parameters were measured simultaneously using two methods: an invasive Swan-Ganz catheterization and a thermodilution technique; and a non-invasive recording using a device for finger arterial pressure waveform analysis. The following parameters were analysed: stroke volume (SV, ml), cardiac output (CO, l/min) and systemic vascular resistance (SVR, dyne × s/cm⁵). The concordance between these two methods was assessed using the variability coefficient calculated according to a Bland-Altman method. Comparing the invasive and non-invasive measures, variability coefficients were: 13, 18 and 11% for SV, SVR and CO, respectively. Similar variability coefficients were obtained when invasive and non-invasive measurements were compared in prespecified subgroups of patients, distinguished based on the presence of decompensation, atrial fibrillation and values of SBP. CONCLUSIONS: Evaluation of haemodynamic parameters using a non-invasive method based on a pressure pulse contour model reveals an adequate concordance with the measures obtained using an invasive approach. Our results suggest that a non-invasive method for haemodynamic monitoring could be applied in clinical practice in patients with advanced heart failure.
No takes yet. Share an insight, caveat, or question.
Sokolski et al. (2011) conducted an observational in advanced heart failure (n=25). Non-invasive recording using a device for finger arterial pressure waveform analysis vs. Invasive Swan-Ganz catheterization and thermodilution technique was evaluated on Concordance between invasive and non-invasive measurements of stroke volume, systemic vascular resistance, and cardiac output (Variability coefficients: 13% for SV, 18% for SVR, 11% for CO). Non-invasive haemodynamic measurement using finger arterial pressure waveform analysis showed adequate concordance with invasive Swan-Ganz catheterization (variability coefficients 11-18%).