Key result
Transcatheter implantation of the CardioMEMS HF System was successfully achieved in 100% (14 of 14) of adult patients with complex congenital heart disease and advanced heart failure.
Why the study?
Heart failure is a common late cardiovascular complication in adult congenital heart disease, necessitating a better understanding of implantable hemodynamic monitor application in this population.
Is transcatheter implantation of an implantable hemodynamic monitor feasible and safe in adults with complex palliated congenital heart disease and advanced heart failure?
Observational (n=14)
Is transcatheter implantation of an implantable hemodynamic monitor feasible and safe in adults with complex palliated congenital heart disease and advanced heart failure?
Transcatheter implantation of an implantable pulmonary artery pressure monitor is technically feasible in adult patients with complex congenital heart disease and advanced heart failure.
OBJECTIVE: We describe the technical considerations of transcatheter implantation of the CardioMEMS™ HF System (Abbott, Abbott Park, IL) in adult patients with complex palliated congenital heart disease (CHD) and advanced heart failure (HF). BACKGROUND: Ambulatory pulmonary artery (PA) pressure monitoring with implantable hemodynamic monitors (IHMs) has been shown to reduce HF-related hospital admissions in non-CHD populations. HF is a common late cardiovascular complication in adult CHD necessitating better understanding of IHM application in this population. METHODS: We analyzed adults with complex CHD and advanced HF who were referred for CardioMEMS™ device implantation (2015-2018). Feasibility of device implantation, defined by successful device implantation and calibration, and procedural outcomes were evaluated. RESULTS: CardioMEMS™ was successfully implanted in all 14 adults (35.5 ± 9.2 years old, 72 ± 12 kg) with complex CHD (single ventricle/Fontan, n = 8 [57%]; d-transposition of the great arteries/atrial switch, n = 6 [43%]). The device was delivered via femoral venous access in 13 (93%) patients and implanted in the left PA in 12 (86%). A long sheath was used in 8 (57%) patients, including 5/6 with an atrial switch operation. There was one device migration that did not require retrieval. CONCLUSIONS: Transcatheter implantation of an IHM is feasible in select complex adult CHD patients with advanced HF. Further studies evaluating integration of ambulatory hemodynamics and the impact on clinical care are needed. This technology has the potential to improve medical management of advanced HF in patients with Fontan and atrial switch physiologies and provide new insights into their ambulatory hemodynamics.
No takes yet. Share an insight, caveat, or question.
Salavitabar et al. (2019) conducted an observational in complex palliated congenital heart disease and advanced heart failure (n=14). CardioMEMS HF System was evaluated on Feasibility of device implantation (successful device implantation and calibration). Transcatheter implantation of the CardioMEMS HF System was successfully achieved in 100% (14 of 14) of adult patients with complex congenital heart disease and advanced heart failure.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: