Key result
Rapid quantitative measurement of troponin T and myoglobin achieved a 66.7% positive diagnostic rate within 3 hours of symptom onset, compared to 50.0% for CK-MB and 16.7% for qualitative troponin T.
Why the study?
Does rapid quantitative measurement of myoglobin and troponin T improve early diagnosis of acute myocardial infarction compared to qualitative troponin T and CK-MB?
Observational (n=34)
Yes
Does rapid quantitative measurement of myoglobin and troponin T improve early diagnosis of acute myocardial infarction compared to qualitative troponin T and CK-MB?
Rapid quantitative bedside measurement of myoglobin and troponin T improves the sensitivity of early AMI diagnosis within 3 hours of symptom onset compared to standard CK-MB or qualitative troponin T testing.
Answers are grounded in this paper and cite their sources.
Bedside troponin testing may accelerate AMI diagnosis; leaves open validation and outcome impact in larger prospective studies.
BACKGROUND: New equipment, the Cardiac Reader(TM), which can measure blood concentrations of troponin T (T) and myoglobin (M) in only 15 min at the bedside was evaluated for early diagnosis of acute myocardial infarction (AMI). METHODS AND RESULTS: A total of 34 consecutive patients with AMI who came to hospital within 24 h after onset were studied. Blood samples were collected from the patients at admission and 6, 12, 24, 48 h after onset to qualitatively and quantitatively measure T, M and creatine kinase-MB fraction. There were 20 patients with positive results by qualitative troponin T test and 29 with positive results by quantitative test. Of the patients who visited hospital within 3 h of onset, 17% were positive by the qualitative test and 67% cases had positive results in the quantitative test. The patients were divided into 2 groups according to the flow grade in the infarct-related coronary artery. In the TIMI 0-1 group (n=28), serum myoglobin concentrations were higher than in the TIMI 3-4 group (n=6) at admission and at their peak. CONCLUSION: The rapid quantitative test of T and M is useful for early diagnosis of AMI and as an indicator of its severity, which can be evaluated from the myoglobin concentration in the hyper-acute phase.
No takes yet. Share an insight, caveat, or question.
Yamamoto et al. (2004) conducted an observational in Acute Myocardial Infarction (n=34). Rapid quantitative measurement of myoglobin and troponin T (Cardiac Reader) vs. CK-MB and qualitative Troponin T was evaluated on Positive diagnostic rate within 3 hours of symptom onset. Rapid quantitative measurement of troponin T and myoglobin achieved a 66.7% positive diagnostic rate within 3 hours of symptom onset, compared to 50.0% for CK-MB and 16.7% for qualitative troponin T.