Key result
Implantation of a novel autologous tissue-engineered heart valve (Biovalve) in 20 adult goats was feasible, showing smooth leaflet movement and characteristic tissue construction up to 3 years 4 months.
Why the study?
To test the feasibility and long-term availability of a novel autologous conduit-type tissue-engineered heart valve without foreign bodies in large animal experiments.
Does a novel autologous tissue-engineered heart valve demonstrate feasibility and long-term function in a large animal model?
Does a novel autologous tissue-engineered heart valve demonstrate feasibility and long-term function in a large animal model?
Answers are grounded in this paper and cite their sources.
A novel autologous tissue-engineered heart valve without foreign bodies demonstrated long-term feasibility, function, and biocompatibility in a large animal model.
Objective: We are developing a novel autologous tissue-engineered heart valve (Biovalve) with a unique in-body tissue engineering. This is expected to be a viable bioprosthesis keeping better biocompatibility. In this study, we developed a conduit-type valve without any foreign bodies and tested the feasibility and long-term availability in large animal experiments. Methods: We created plastic molds for Biovalves with 3D printer easily and quickly considering the recipient character. We embedded them in the subcutaneous spaces of adult goats about 2 months. After extracting the molds with the tissue en-block and removing the plastic molds only, Biovalve with tri-leaflets similar to those of the native valves were constituted from completely autologous connective tissues and fibroblasts. Total 20 conduit-type Biovalves were implanted in the apico-aortic bypass or the pulmonary artery of goats, (8 and 12, respectively). No anticoagulants were used after implantation. Results: Biovalves were successfully implanted and showed smooth movement of the leaflets with a little regurgitation in angiogram, and the maximum duration reached to 3 years 4 months. Histological examination of the Biovalves showed the autologous cells covering the laminar surface of the valve leaflets as the endothelium and also getting inside to construct characteristic tissues like native leaflets. Conclusions: The Biovalves have a potential to be used for viable bioprosthetic valves and to keep better function and biocompatibility longer than current ones.
No takes yet. Share an insight, caveat, or question.
Takewa et al. (2019) studied Heart valve replacement (animal model) (n=20). Autologous tissue-engineered heart valve (Biovalve) was evaluated on Feasibility and long-term availability. Implantation of a novel autologous tissue-engineered heart valve (Biovalve) in 20 adult goats was feasible, showing smooth leaflet movement and characteristic tissue construction up to 3 years 4 months.