Does eject speed profiling reduce peak internal pressures in the MagScrew total artificial heart compared to unprofiled eject velocity in an in-vitro model?
Does eject speed profiling reduce peak internal pressures in the MagScrew total artificial heart compared to unprofiled eject velocity in an in-vitro model?
In-vitro testing of the MagScrew total artificial heart demonstrates that eject speed profiling can reduce peak internal pump pressures, which may help prevent premature bioprosthetic valve failure.
Answers are grounded in this paper and cite their sources.
May reduce peak pressures in MagScrew TAH; hypothesis-generating for valve durability, needs in-vivo confirmation.
Purpose: Incompetent inflow valves have been reported with clinical pulsatile LVADs that use bioprosthetic valves. Absolute pressures and/or instantaneous pressure changes within these devices have been suspected to cause premature valve failure. The MagScrew total artificial heart (TAH) was evaluated to quantify absolute pump internal pressures and instantaneous pressure changes under different operating conditions. Methods: The MagScrew TAH is a passively filling pulsatile pump which utilizes a reciprocating magnetic actuating mechanism to propel blood into the circulation. Multiple pump eject parameters were varied which yielded various speed profiles. A total of 49 speed profiles were tested for the left eject phase, and 38 speed profiles were tested for the right eject phase. These were evaluated in-vitro under the following operating conditions: 220BPM, 100% pump fill, a mean aortic pressure of 100 mmHg and a mean pulmonary artery pressure of 20 mmHg. The pressure inside the left and right pump chambers was measured with a Millar Mikro-Tip catheter and captured using Power Lab at a rate of 40 kHz. Results: The pump chamber peak pressures operating with unprofiled eject velocity measured 183–226 mmHg for the left and 126–130 mmHg for the right. Eject speed profiling for both pumps reduced the peak pressure by as much as 25 and 30 mmHg for the left and right pump, respectively. Future studies will look into software controlled optimization of the eject speed profiles under any operating condition.
No takes yet. Share an insight, caveat, or question.
Flick et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: