The MagScrew Total Artificial Heart demonstrates effective physiologic responses to preload and afterload in an in vitro mock circulatory loop, supporting its potential for maintaining appropriate perfusion and preventing pulmonary congestion.
The purpose of this study was to evaluate the in vitro responses to preload and afterload of our total artificial heart (TAH), the MagScrew TAH. The TAH consists of two blood pumps and a control logic, developed at the Cleveland Clinic, OH, and the MagScrew actuator and its electronic control system, developed by Foster-Miller Technologies, Inc., Albany, NY. Tests were performed on a mock circulatory loop, using water as a test fluid. Preload sensitivity of the Mag-Screw TAH demonstrated a Frank-Starling response to preload in automatic mode. A peak flow of 10 L/min was obtained, with a left atrial pressure of 13 mm Hg. The relationship between right atrial pressure and left atrial pressure was well balanced when tested with a left bronchial shunt flow of 5% and a range of pulmonary artery and aortic pressures. With respect to afterload response, the left pump showed a relatively low sensitivity, which allowed the pump to maintain perfusion over a wide range of aortic pressures. The right pump, on the other hand, was much more sensitive to pulmonary artery pressure, which provided a measure of protection against pulmonary congestion. The very effective physiologic response of the MagScrew TAH is believed to result from employment of a left master, alternating ejection control logic, high inherent sensitivity of the blood pumps to atrial pressure, a lower effective stroke volume for the right pump, and a scaling of right side motor ejection voltage to 80% of that used for the left side ejection.
No takes yet. Share an insight, caveat, or question.
Weber et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: