A novel pneumatic filtering method was developed and experimentally validated to reduce body motion artifacts in a noninvasive pneumatic biomeasurement system.
Answers are grounded in this paper and cite their sources.
May improve ambulatory signal fidelity in pneumatic systems; leaves open clinical validation before practice adoption.
The pneumatic biomeasurement system is a noninvasive, constraint-free method for monitoring the heartbeat, respiration and other vital signs of a person in bed. Body microvibrations transmitted through the bed cushion act on the air in a thin air mattress under the cushion. Changes in air pressure synchronized with the microvibrations are detected by a pressure sensor. However, the high sensitivity of the sensor results in output saturated with body motion artifacts. This paper describes a novel method for pneumatic filtering of these artifacts. A theoretical model was developed as the basis for designing a pneumatic filter to reduce body motion artifacts. The validity of the proposed high-pass filtering method was examined through experiments.