Key result
In vitro testing of four 22 mm polymeric valved conduits demonstrated successful balloon dilation to 23.26 ± 0.38 mm with increased effective orifice areas and maintained low regurgitation.
Why the study?
Currently available heart valve prostheses lack growth potential, subjecting pediatric patients with heart valve disease to multiple valve replacement surgeries with compounding risks.
An in vitro proof-of-concept study demonstrates the feasibility of a polymeric balloon-expandable valved conduit that can be dilated to accommodate growth, potentially reducing reoperations in pediatric patients.
Currently available heart valve prostheses have no growth potential, requiring children with heart valve diseases to endure multiple valve replacement surgeries with compounding risks. This study demonstrates the in vitro proof of concept of a biostable polymeric trileaflet valved conduit designed for surgical implantation and subsequent expansion via transcatheter balloon dilation to accommodate the growth of pediatric patients and delay or avoid repeated open-heart surgeries. The valved conduit is formed via dip molding using a polydimethylsiloxane-based polyurethane, a biocompatible material shown here to be capable of permanent stretching under mechanical loading. The valve leaflets are designed with an increased coaptation area to preserve valve competence at expanded diameters. Four 22 mm diameter valved conduits are tested in vitro for hydrodynamics, balloon dilated to new permanent diameters of 23.26 ± 0.38 mm, and then tested again. Upon further dilation, two valved conduits sustain leaflet tears, while the two surviving devices reach final diameters of 24.38 ± 0.19 mm. After each successful dilation, the valved conduits show increased effective orifice areas and decreased transvalvular pressure differentials while maintaining low regurgitation. These results demonstrate concept feasibility and motivate further development of a polymeric balloon-expandable device to replace valves in children and avoid reoperations.
No takes yet. Share an insight, caveat, or question.
Li et al. (2023) studied Pediatric heart valve disease (n=4). Biostable polymeric trileaflet valved conduit was evaluated on Hydrodynamics and balloon dilation capability. In vitro testing of four 22 mm polymeric valved conduits demonstrated successful balloon dilation to 23.26 ± 0.38 mm with increased effective orifice areas and maintained low regurgitation.