Why the study?
Device-based algorithms offer automated CRT optimization, but their clinical adoption is often ad-hoc. The authors evaluated how QuickOpt-guided VV delay programming performs against the heuristic of QRS complex narrowing.
Does QuickOpt-guided VV delay programming reduce QRS complex duration compared to synchronous biventricular pacing in patients with CRT?
Observational (n=37)
Yes
Does QuickOpt-guided VV delay programming reduce QRS complex duration compared to synchronous biventricular pacing in patients with CRT?
The QuickOpt algorithm systematically favors left-preexcitation VV delay, leading to inadvertent QRS prolongation compared to synchronous biventricular pacing.
Answers are grounded in this paper and cite their sources.
QuickOpt VV optimization may prolong QRS in CRT; leaves open need for QRS-validated algorithms in prospective trials.
BACKGROUND: Device-based algorithms offer the potential for automated optimization of cardiac resynchronization therapy (CRT), but the process for accepting them into clinical use is currently still ad-hoc, rather than based on pre-clinical and clinical testing of specific features of validity. We investigated how the QuickOpt-guided VV delay (VVD) programming performs against the clinical and engineering heuristic of QRS complex shortening by CRT. METHODS: A prospective, 2-center study enrolled 37 consecutive patients with CRT. QRS complex duration (QRSd) was assessed during intrinsic atrioventricular conduction, synchronous biventricular pacing, and biventricular pacing with QuickOpt-proposed VVD. The measurements were done manually by electronic calipers in signal-averaged and magnified 12-lead QRS complexes. RESULTS: Native QRSd was 174 ± 22 ms. Biventricular pacing with empiric AVD and synchronous VVD resulted in QRSd 156 ± 20 ms, a significant narrowing from the baseline QRSd by 17 ± 27 ms, P = 0.0003. In 36 of 37 patients, the QuickOpt algorithm recommended left ventricular preexcitation with VVD of 42 ± 18 ms (median 40 ms; interquartile range 30-55 ms, P <0.00001). QRSd in biventricular pacing with QuickOpt-based VVD was significantly longer compared with synchronous biventricular pacing (168 ± 25 ms vs. 156 ± 20 ms; difference 12 ± 11ms; P <0.00001). This prolongation correlated with the absolute VVD value (R = 0.66, P <0.00001). CONCLUSIONS: QuickOpt algorithm systematically favours a left-preexcitation VVD which translates into a significant prolongation of the QRSd compared to synchronous biventricular pacing. There is no reason to believe that a manipulation that systematically widens QRSd should be considered to optimize physiology. Device-based CRT optimization algorithms should undergo systematic mechanistic pre-clinical evaluation in various scenarios before they are tested in large clinical studies.
No takes yet. Share an insight, caveat, or question.
Sedláček et al. (2022) conducted an observational in Cardiac resynchronization therapy (n=37). QuickOpt-guided VV delay (VVD) programming vs. Synchronous biventricular pacing was evaluated on QRS complex duration (QRSd) (MD 12, p=<0.00001).