Why the study?
Does coating defibrillation leads with ePTFE or backfilling with medical adhesive improve extraction outcomes and reduce tissue trauma in a sheep model?
Does coating defibrillation leads with ePTFE or backfilling with medical adhesive improve extraction outcomes and reduce tissue trauma in a sheep model?
Answers are grounded in this paper and cite their sources.
Coating defibrillation shocking coils with ePTFE or backfilling with medical adhesive reduces tissue ingrowth and facilitates transvenous lead removal with less tissue trauma in an animal model.
BACKGROUND: Permanent leads with shocking coils for defibrillation therapy are sometimes implanted in the coronary sinus (CS) and great cardiac vein (GCV). These shocking coils, as documented by pathologic examination of animal investigations, often become tightly encapsulated by fibrosis and can be very difficult to remove. METHODS: One of three configurations of the Guidant model 7109 Perimeter coronary sinus shocking lead was implanted into the distal portion of the GCV of 24 sheep for up to 14 months. Group 1 had unmodified coils (control), group 2 had coils backfilled with medical adhesive (MA), and Group 3 had coils coated with expanded polytetrafluoroethylene (ePTFE). Eighteen leads, three from each group at 6 and 14 months were transvenously extracted from the left jugular vein. The remaining six animals were not subject to extraction. All animals were euthanized for pathological and microscopic examination. RESULTS: All six of the control, three of the MA, and one of the ePTFE leads required the use of an electrosurgical dissection sheath (EDS) for extraction. Five control, two MA, and none of the ePTFE leads had significant fibrotic attachments to the shocking coils. Significant trauma was observed at necropsy for those leads requiring the use of the EDS for extraction. CONCLUSIONS: Tissue ingrowth is a major impediment to the removal of defibrillation leads implanted in the CS and GCV of sheep. Reduction of tissue ingrowth by coating the shocking coils with ePTFE or by backfilling with MA facilitates transvenous lead removal with reduced tissue trauma.
No takes yet. Share an insight, caveat, or question.
Wilkoff et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: