Answers are grounded in this paper and cite their sources.
Peak extractor circuits are an important link in most pulse-height analysis systems. A novel architecture with performance exceeding previous state-of-the-art designs is presented. A droop rate ranging from 125 mu V/ mu s and a charge injection of 5 mV are reported along with a 2.2 MHz bandwidth, a 38 V/ mu s slew rate, and integral nonlinearity of 50 PPM. The low droop rate allows temporary storage of the signal in an analog FIFO (first-in first-out) with minimal peak broadening, suitable for high-resolution low-energy X-ray analysis. This configuration has been implemented in an analog-to-digital converter system using the peak-detect and hold circuit architecture. The complete system had a differential linearity of 1.5% at 64 K conversion gain, an integral linearity of +or-0.016%, and an effective conversion time of 1.9 mu s for a count rate of 210000 C/s to the input of a pulse processor producing 3- mu s-wide pulses.>
No takes yet. Share an insight, caveat, or question.
Buckens et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: