Key result
An OCT-integrated ablation catheter demonstrated accuracies of 99.96% and 92.88% for discerning catheter contact and orientation in swine left-atrial specimens.
Why the study?
Catheter contact and orientation critically impact lesion formation in atrial fibrillation ablation, but must currently be inferred indirectly.
Does OCT-integrated ablation catheter imaging accurately assess catheter contact and orientation in swine left-atrial specimens?
Does OCT-integrated ablation catheter imaging accurately assess catheter contact and orientation in swine left-atrial specimens?
OCT M-mode imaging integrated into ablation catheters can highly accurately assess catheter contact and orientation in preclinical models, potentially improving radiofrequency ablation for atrial fibrillation.
May enable precise real-time catheter assessment in AF ablation; leaves open in vivo and clinical validation.
The efficacy of catheter ablation treatment for atrial fibrillation is directly impacted by the quality of lesion formation. Two parameters that are critical for maximizing energy delivery are sustained catheter contact and orientation. Currently, these parameters must be inferred indirectly through tactile feedback or measurements of bioelectrical impedance and tip force. In this work, we propose a method for discerning contact and orientation based on direct endomyocardial imaging mediated by optical coherence tomography (OCT)-integrated ablation catheters. A two-stage classifier is developed to deduce contact parameters from M-mode images. Experimental validation within swine left-atrial specimens demonstrate accuracies of 99.96% and 92.88% for contact and orientation stages, respectively. These results highlight the potential of OCT M-mode imaging for guiding catheter placement during radiofrequency ablation interventions.
No takes yet. Share an insight, caveat, or question.
Yu et al. (2019) studied Atrial fibrillation. Optical coherence tomography (OCT)-integrated ablation catheters was evaluated on Accuracy of discerning catheter contact and orientation. An OCT-integrated ablation catheter demonstrated accuracies of 99.96% and 92.88% for discerning catheter contact and orientation in swine left-atrial specimens.