Key result
A prototyped forward-imaging OCT catheter (2-mm scan diameter, 30-μm spot size) successfully demonstrated real-time monitoring of cardiac radio-frequency ablation lesion formation ex vivo.
Why the study?
Does an OCT forward-imaging catheter allow real-time monitoring of cardiac radio-frequency ablation lesion formation in an ex vivo swine model?
Does an OCT forward-imaging catheter allow real-time monitoring of cardiac radio-frequency ablation lesion formation in an ex vivo swine model?
A prototyped OCT forward-imaging catheter successfully demonstrated real-time monitoring of cardiac radio-frequency ablation lesion formation in an ex vivo swine model.
Hypothesis-generating for OCT-guided ablation monitoring; in vivo human studies required before clinical adoption.
Radio-frequency ablation (rfa) is the standard of care for the treatment of cardiac arrhythmias; however, there are no direct measures of the successful delivery of ablation lesions. Optical coherence tomography (OCT) imaging has the potential to provide real-time monitoring of cardiac rfa therapy, visualizing lesion formation and assessing tissue contact in the presence of blood. A rfa-compatible forward-imaging conical scanning probe is prototyped to meet this need. The forward-imaging probe provides circular scanning, with a 2-mm scan diameter and 30-mum spot size. During the application of rf energy, dynamics are recorded at 20 frames per second with a 40-kHz A-line rate. Real-time monitoring of cardiac rfa lesion formation and imaging in the presence of blood is demonstrated ex vivo in a swine left ventricle with a forward, flexible, circular scanning OCT catheter.
No takes yet. Share an insight, caveat, or question.
Fleming et al. (2010) studied cardiac arrhythmias. Optical coherence tomography (OCT) forward-imaging catheter was evaluated on Real-time monitoring of cardiac rfa lesion formation. A prototyped forward-imaging OCT catheter (2-mm scan diameter, 30-μm spot size) successfully demonstrated real-time monitoring of cardiac radio-frequency ablation lesion formation ex vivo.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: