Why the study?
Does a linear correction method resolve the discrepancy between whole-blood and fresh plasma ISI calibrations for POCT PT monitors?
Does a linear correction method resolve the discrepancy between whole-blood and fresh plasma ISI calibrations for POCT PT monitors?
A linear correction method resolves the discrepancy between whole-blood and fresh plasma ISI calibrations for TAS PT-NC test cards, simplifying POCT PT monitor calibration.
Answers are grounded in this paper and cite their sources.
Linear correction may simplify POCT PT monitor calibration; leaves open prospective validation before clinical adoption.
BACKGROUND: The International Sensitivity Index (ISI) calibration of whole-blood prothrombin time (PT) monitors for point-of-care testing (POCT) described by Tripodi et al. (Thromb Haemost 1993;70:921-4) has been shown to be dependable but is too complex and demanding. The use of plasma would simplify calibration of whole-blood POCT PT monitors, but important differences may exist between the ISI for whole blood and plasma calibrations. METHODS: In a 10-center calibration study of two POCT whole-blood monitoring systems (CoaguChek Mini and TAS PT-NC), we characterized the relationship between the log PT for whole blood and fresh plasma with use of single lots of test strips/cards. This relationship (linear) was used to correct the difference between the whole-blood and plasma ISI. The reliability of the correction with different lots of test strips/cards was assessed at three centers. The linear relationship was used to correct the difference in the whole-blood and plasma ISI with four other lots of TAS PT-NC cards and with two additional lots of CoaguChek Mini test strips. RESULTS: The correction decreased the ISI difference from 13.3% to 0.9% for the TAS PT-NC and from 5.7% to 0.6% for the CoaguChek Mini. In assessments at three centers, which included different lots of test strips/cards, the mean ISI difference was markedly decreased with the TAS PT-NC but not with the CoaguChek Mini, for which the mean ISI difference increased slightly. CONCLUSIONS: The proposed correction resolves the discrepancy between whole-blood and fresh plasma ISI calibrations with TAS PT-NC test cards. The CoaguChek Mini systems could be calibrated without this correction.
No takes yet. Share an insight, caveat, or question.
Poller et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: