Why the study?
Can the ProTime near-patient testing monitor be calibrated according to the WHO international sensitivity index (ISI) model?
Can the ProTime near-patient testing monitor be calibrated according to the WHO international sensitivity index (ISI) model?
The ProTime near-patient monitor for prothrombin time testing demonstrates acceptable ISI calibration and INR agreement with reference standards, supporting its potential for patient self-testing.
Answers are grounded in this paper and cite their sources.
Supports ProTime calibration per WHO model; leaves open clinical adoption for self-testing pending outcome studies.
Near-patient testing devices (monitors) capable of measuring prothrombin time on an unmeasured drop of blood would be suitable alternatives to centralized laboratory monitoring of patients receiving oral anticoagulants. The essential prerequisite for the use of these monitors is their conformity to the international sensitivity index (ISI) calibration model recommended by the World Health Organization. We report on the ISI calibration of the ProTime monitor (International Technidyne, Edison, NJ) designed and approved for patient self-testing. According to our results, this monitor can be calibrated by adopting the model already used for other monitors. The apparent ISI was close to unity. Overall, the international normalized ratio values displayed by the monitor agreed with those measured with the international reference preparation for thromboplastin. Confirmation of these results in a large multicenter study is warranted.
No takes yet. Share an insight, caveat, or question.
Bressi et al. (2003) studied this question.