Why the study?
Can the ProTime monitor be calibrated according to the WHO international sensitivity index model for measuring prothrombin time?
Can the ProTime monitor be calibrated according to the WHO international sensitivity index model for measuring prothrombin time?
The ProTime monitor for near-patient prothrombin time testing can be accurately calibrated to WHO standards, supporting its potential use for patient self-testing.
Answers are grounded in this paper and cite their sources.
Supports ProTime calibration for near-patient INR; leaves open outcome validation before self-testing adoption.
Near-patient testing devices (monitors) capable of measuring prothrombin time on an unmeasured drop of blood would be suitable alternatives to centralized laboratory monitoring of patients receiving oral anticoagulants. The essential prerequisite for the use of these monitors is their conformity to the international sensitivity index (ISI) calibration model recommended by the World Health Organization. We report on the ISI calibration of the ProTime monitor (International Technidyne, Edison, NJ) designed and approved for patient self-testing. According to our results, this monitor can be calibrated by adopting the model already used for other monitors. The apparent ISI was close to unity. Overall, the international normalized ratio values displayed by the monitor agreed with those measured with the international reference preparation for thromboplastin. Confirmation of these results in a large multicenter study is warranted.