Key result
Preprocedural CT-SQUEEZ-derived target selection for LV lead placement produced a similar mean acute hemodynamic response compared to the best achievable response (20.4% vs 24.9%; P=0.36).
Why the study?
Does preimplant cardiac CT with the SQUEEZ algorithm effectively identify optimal venous targets for LV lead placement in patients undergoing CRT upgrade?
Observational (n=18)
Does preimplant cardiac CT with the SQUEEZ algorithm effectively identify optimal venous targets for LV lead placement in patients undergoing CRT upgrade?
Answers are grounded in this paper and cite their sources.
Preprocedural cardiac CT using the SQUEEZ algorithm can effectively identify optimal venous targets for LV lead placement in CRT upgrades, yielding hemodynamic responses comparable to the best achievable.
BACKGROUND: Optimal lead positioning is an important determinant of cardiac resynchronization therapy (CRT) response. OBJECTIVE: The purpose of this study was to evaluate cardiac computed tomography (CT) selection of the optimal epicardial vein for left ventricular (LV) lead placement by targeting regions of late mechanical activation and avoiding myocardial scar. METHODS: Eighteen patients undergoing CRT upgrade with existing pacing systems underwent preimplant electrocardiogram-gated cardiac CT to assess wall thickness, hypoperfusion, late mechanical activation, and regions of myocardial scar by the derivation of the stretch quantifier for endocardial engraved zones (SQUEEZ) algorithm. Cardiac venous anatomy was mapped to individualized American Heart Association (AHA) bull's-eye plots to identify the optimal venous target and compared with acute hemodynamic response (AHR) in each coronary venous target using an LV pressure wire. RESULTS: Fifteen data sets were evaluable. CT-SQUEEZ-derived targets produced a similar mean AHR compared with the best achievable AHR (20.4% ± 13.7% vs 24.9% ± 11.1%; P = .36). SQUEEZ-derived guidance produced a positive AHR in 92% of target segments, and pacing in a CT-SQUEEZ target vein produced a greater clinical response rate vs nontarget segments (90% vs 60%). CONCLUSION: Preprocedural CT-SQUEEZ-derived target selection may be a valuable tool to predict the optimal venous site for LV lead placement in patients undergoing CRT upgrade.
No takes yet. Share an insight, caveat, or question.
Behar et al. (2017) conducted an observational in Patients undergoing cardiac resynchronization therapy (CRT) upgrade (n=18). Preimplant electrocardiogram-gated cardiac CT with SQUEEZ algorithm vs. Best achievable acute hemodynamic response was evaluated on Mean acute hemodynamic response (AHR) (p=.36). Preprocedural CT-SQUEEZ-derived target selection for LV lead placement produced a similar mean acute hemodynamic response compared to the best achievable response (20.4% vs 24.9%; P=0.36).
Synapse has enriched 10 related papers. Consider them for comparative context: