The Immulite 2000 cTnI assay has a 99th percentile limit of 0.32 μg/L, but its precision at low concentrations is unsatisfactory, necessitating a cutoff of 0.4 μg/L for a CV of 12% in routine clinical practice.
Answers are grounded in this paper and cite their sources.
May require adjusted cutoff for acceptable precision in clinical use; leaves open assay standardization needs.
Recently, the Joint European Society of Cardiology/American College of Cardiology committee for the redefinition of myocardial infarction proposed that “any amount of myocardial necrosis caused by ischemia should be labeled as an infarct” (1). The same committee agreed that a troponin concentration higher than the 99th percentile of a reference control group is one of the most specific (biochemical) markers of myocardial infarction. The maximum allowable imprecision at this decision limit has arbitrarily been set at a CV of 10% (1)(2) although it is not stated whether the 10% relates to total CV. Because an increasing number of nonstandardized commercial methods for the determination of cardiac troponin I (cTnI) are available, the analytical characteristics of every assay need to be established and published (1)(3). We describe the characteristics of the Immulite 2000 cTnI assay manufactured by Diagnostic Products Corporation. The Immulite 2000 cTnI sandwich assay uses a monoclonal binding antibody and an alkaline phosphatase-conjugated polyclonal tracer antibody. Both antibodies recognize epitopes between amino acids 33 and 110 of cTnI. After incubation, excess conjugate is removed and chemiluminescent substrate is added. Chemiluminescence is proportional to the concentration of cTnI in the sample. We assessed the detection limit of the assay by measuring 20 times a cTnI-free calibrator supplied by DPC. Each result was <0.01 μg/L. Therefore, the detection limit, defined as the signal 2 SD above the mean of the analyte-free calibrator, is <0.02 μg/L. To determine the 99th percentile of a control group, we analyzed heparinized plasma from 315 healthy volunteers from the outpatient clinic (approximately equal numbers of males and females, almost exclusively Caucasians). A total of 308 samples (98%) had cTnI concentrations <0.01 μg/L, and 7 samples contained 0.03–0.33 μg/L cTnI. The 99% limit was calculated to be 0.32 μg/L. The authors of another study, with 117 participants, reported the 99% limit, measured in serum on the Immulite analyzer, as 0.48 μg/L (4). Immulite serum cTnI values are 5–10% higher than those measured in heparin plasma (4)(5). In accordance with the NCCLS EP5-T2 guideline (6), we analyzed five pools of heparinized plasma in duplicate once a day over 20 days and constructed an imprecision profile in the lower assay range. Our results (Fig. 1) are similar to those described previously for the Immulite analyzer (4)(5). A total CV of 10% was reached at a concentration of 0.7 μg/L. However, considering the 99th percentile found in the reference population, we arbitrarily chose a cutoff value of 0.4 μg/L at a CV of 12% for routine clinical practice to increase the sensitivity of the assay for detection of minimal cardiac damage. Precision profile of the Immulite 2000 cTnI assay. In summary, the Immulite 2000 cTnI assay is analytically suited for the triage of patients presenting with symptoms of myocardial damage. Like many other commercial troponin assays (7), its precision at low troponin concentrations is unsatisfactory, so future generations of the assay need to have improved analytical sensitivity. We thank DPC Nederland (Breda, The Netherlands) for the cTnI reagent sets used in this study.
No takes yet. Share an insight, caveat, or question.
Scharnhorst et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: