Why the study?
Does the presence of retained CIED fragments after heart transplantation increase the risk of adverse outcomes?
Does the presence of retained CIED fragments after heart transplantation increase the risk of adverse outcomes?
Retained CIED fragments are common after heart transplantation and are associated with an increased risk of upper extremity deep venous thrombosis and longer radiation exposure during biopsies.
Answers are grounded in this paper and cite their sources.
Retained CIED fragments may increase post-transplant UEDVT risk; leaves open whether extraction improves outcomes.
We performed a retrospective review of 402 consecutive patients who underwent heart transplantation at our institution between January 2009 and March 2017. A retained cardiovascular implantable electronic device (CIED) fragment was identified after transplantation in 49 of the 301 patients (16.2%) with CIED at baseline. Patients with retained fragments had leads with longer dwell times (median 2596 [1982, 3389] vs 1384 [610, 2202] days, P < .001), higher prevalence of previously abandoned leads (14.3% vs 2.8%, P = .003), and dual-coil defibrillator leads (98% vs 81%, P = .001) compared with patients without retained fragments. Five patients (10%) with retained CIED fragments underwent magnetic resonance imaging without adverse events. There was no difference in overall mortality between patients with and without CIED fragments (12% vs 11%, P = .81) Patients with retained fragments located in the superior vena cava had significantly higher fluoroscopic times (3.3 vs 2.9 minutes, P = .024) during subsequent endomyocardial biopsies. In a competing risk analysis, presence of a retained CIED fragment was associated with upper extremity deep venous thrombosis (sub hazard ratio [HR] 2.19, 95% confidence interval [CI] 1.17-4.10, P = .014) but not bloodstream infection after adjusting for potential confounders. In summary, retained CIED fragments are common after heart transplantation, and are associated with longer radiation exposure during biopsy procedures and upper extremity deep venous thrombosis.
No takes yet. Share an insight, caveat, or question.
Álvarez et al. (2018) studied this question.