Highlights a complex mechanism of inappropriate ICD shocks involving far-field atrial oversensing exacerbated by structural and electrical changes in giant-cell myocarditis.
Clinicians should evaluate lead slack and atrial arrhythmias in unexplained ICD shocks with myocarditis; leaves open optimal prevention strategies in complex device cases.
We report a unique case of delivery of inappropriate implantable cardioverter-defibrillator therapies related to a “perfect storm”: presence of an integrated lead, insufficient lead slack related to right heart dilation resulting in shock coil misplacement, myocarditis with loss of R waves, and the concomitant occurrence of an incessant atrial tachycardia. (Level of Difficulty: Advanced.)
No takes yet. Share an insight, caveat, or question.
Datta et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: