Key result
CMR provided the most accurate aortic annulus measurements in vitro, while MSCT and 3DE were reasonable in vivo alternatives that slightly overestimated and underestimated dimensions, respectively.
Why the study?
Does CMR provide more accurate aortic annulus measurements compared to 3DE and MSCT in in vitro and in vivo models?
Observational (n=15)
Does CMR provide more accurate aortic annulus measurements compared to 3DE and MSCT in in vitro and in vivo models?
Answers are grounded in this paper and cite their sources.
CMR provides the most accurate in vitro measurements of the aortic annulus, while MSCT and 3DE are reasonable in vivo alternatives that slightly overestimate and underestimate dimensions, respectively.
OBJECTIVES: To determine the accuracy of calcium-containing rings measurements imaged by three-dimensional echocardiography (3DE), multi-slice CT (MSCT) and cardiac magnetic resonance (CMR) under ideal conditions against the true ring dimensions. To compare the accuracy of aortic annulus (AoA) measurements in ex vivo human hearts using 3DE, MSCT and CMR. To determine the accuracy of AoA measurements in an in vivo human model. DESIGN: 3DE, MSCT and CMR imaging were performed on 30 calcium-containing rings and 28 explanted human hearts. Additionally, 15 human subjects with clinical indication for MSCT underwent 3DE. Two experts in each modality measured the images. MAIN OUTCOME MEASURES: Bias and intraclass correlation coefficient for accuracy of imaging measurements when compared with actual ring dimensions. Bias, intraclass correlation coefficient and variability were obtained: (1) when comparing explanted human heart AoA measurements from the two remaining imaging modalities with the most accurate one as determined from the ring measurements and (2) in in vivo human AoA measurements. Analysis was repeated on explanted heart subgroups divided by aortic valve Agatston score. RESULTS: Against the known ring dimensions, CMR had the highest accuracy and the lowest variability. MSCT measurements had high accuracy but wider variability and 3DE had the lowest accuracy with the largest variability. When 3DE and MSCT were compared with CMR, 3DE underestimated and MSCT overestimated AoA dimensions, but inter-measurement variability of 3DE and MSCT were similar. When divided by Agatston score, both 3DE and MSCT measurements were larger and showed greater variability with increasing calcium burden. The in vivo study showed that the correlation between 3DE and MSCT measurements was high; however, 3DE measurements were smaller than those measured with MSCT. CONCLUSIONS: In the in vitro model, CMR measurements were the most accurate for assessing the actual dimensions suggesting that further investigations on its role in AoA measurement in TAVR are needed. However from the in vivo model, MSCT and 3DE are reasonable alternatives with the understanding that they can slightly overestimate and underestimate annular dimensions, respectively.
No takes yet. Share an insight, caveat, or question.
Tsang et al. (2012) conducted an observational in Clinical indication for MSCT (n=15). 3DE, MSCT, and CMR vs. True ring dimensions / CMR was evaluated on Bias and intraclass correlation coefficient for accuracy of imaging measurements. CMR provided the most accurate aortic annulus measurements in vitro, while MSCT and 3DE were reasonable in vivo alternatives that slightly overestimated and underestimated dimensions, respectively.