Answers are grounded in this paper and cite their sources.
To better understand the performance of the SP3000 time-of-flight PET (positron emission tomography), machine, the authors performed Monte Carlo calculations of photon interactions in a BaF/sub 2/ detector module and calculations of crystal optics for BaF/sub 2/ UV scintillation light. Evaluated were photon detection sensitivities, the amount of photon spillover, crystal crosstalk, interring miscodes, and true-to-random coincidence ratios as a function of the lower-level energy discriminator setting and crystal/photomultiplier tube energy resolution. UV light collection efficiencies for the timing tube and decode tube were calculated assuming different crystal geometries, reflective materials, and reflectances. For a typical SP3000 detector module the detection efficiency for single 511-keV photons varied between 20 and 60% depending on the angle of incidence.>
No takes yet. Share an insight, caveat, or question.
Bice et al. (1990) studied this question.