Why the study?
Long-term durability is critical as TAV replacement extends to younger and lower-risk patients, and increased leaflet mechanical stress is a primary determinant of valve structural deterioration.
How do valve size, stroke volume, and degree of oversizing affect leaflet bending stress in self-expanding transcatheter aortic valves in an in vitro model?
How do valve size, stroke volume, and degree of oversizing affect leaflet bending stress in self-expanding transcatheter aortic valves in an in vitro model?
In vitro testing demonstrates that moderate transcatheter aortic valve oversizing increases leaflet bending stress, particularly in small annuli, potentially compromising long-term durability.
Answers are grounded in this paper and cite their sources.
Moderate oversizing was associated with higher leaflet bending stress in small annuli; leaves open effects on long-term TAV durability.
Transcatheter aortic valve (TAV) replacement has become a viable alternative to surgery for high and intermediate risk patients with severe aortic stenosis. This technology may extend to the younger and lower risk patients. In this population, long-term durability of the TAV is key. Increased leaflet mechanical stress is one of the main determinants of valve structural deterioration. This in vitro study aims at evaluating leaflet bending stress (LBS) in the self-expanding TAV for different valve sizes, stroke volumes (SV), and degrees of valve oversizing (OS). Three different sizes (23, 26, and 29 mm) of CoreValve (CV) were tested on a pulse duplicator in annulus size ranging from 17 to 26 mm. Leaflet bending stress and bending of the leaflet coaptation line in diastole pinwheeling index (PI) were measured using high-speed camera imaging (1000 images/s). For each given CV and annulus size, geometric orifice area (GOA) increased significantly with OS (P < .001) and SV (P = .001). LBS decreased with increasing prosthesis size and aortic annulus (AA) size while increasing with SV (P < .03). The largest value of peak LBS (3.79 MPa) was obtained with the CV 23 mm in AA of 17 mm (%OS = 35%), SV 90 mL and the smallest value (0.99 MPa) for the CV 29 mm in AA of 26 mm (%OS = 12%), SV 30 mL. On multivariable analysis, LBS increased independently with larger OS, smaller AA size and higher SV. The PI increased with decreasing AA size and increasing OS. Moderate valve OS, such as generally used for transcatheter aortic valve implantation, is associated with increased LBS during valve opening and closing, especially in small annuli. Hence, TAV OS may negatively impact long-term valve durability.
No takes yet. Share an insight, caveat, or question.
Stanová et al. (2020) studied this question.