The poor performance of current parallel-hole collimators is an impediment to planar high-resolution gamma-ray imaging, even when high-resolution semiconductor detector arrays are available. High-resolution parallel-hole collimators are possible but have not been fabricated because conventional collimator construction techniques severely limit achievable bore size and septal thickness. We describe development and testing of a high-resolution collimator with 4096 260-/spl mu/m square bores and 380 /spl mu/-m pitch, matched to our existing 2.5 cm/spl times/2.5 cm hybrid 64/spl times/64 CdZnTe arrays with multiplexer readout. The collimator is a laminar composite of about 100 layers of W sheets produced by photolithography and has an efficiency of 5/spl times/10/sup -5/. We have demonstrated sub-millimeter spatial resolution at 140 keV in both phantom and animal imaging using this system. We present images of resolved individual vertebrae in the spine of a mouse and lymphatic channels and nodes in a rat. The collimator and semiconductor array could form a compact module for use in a wide variety of gamma-ray imaging systems.
No takes yet. Share an insight, caveat, or question.
Kastis et al. (2000) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: