Fluid penetration in an ICD connector port can cause electrical artifacts leading to inappropriate shock delivery.
Answers are grounded in this paper and cite their sources.
Requires vigilance for connector artifacts during ICD troubleshooting; case report leaves open incidence and preventive strategies.
We report a unique case of fluid penetration, 3 months after implantation, in the connector port of an automatic implantable cardioverter defibrillator (ICD) with transvenous subcutaneous lead system. The patient had coronary artery disease and recurrent episodes of ventricular fibrillation, the fluid caused electrical signals interpreted as ventricular fibrillation by the device, which triggered shock delivery.