Key result
A technique using two reflective pulse signals from a single green LED switched between high and low light intensities accurately cancelled out motion artifacts under various conditions.
A novel technique using a single green LED switched between high and low intensities can accurately cancel motion artifacts in wearable photoplethysmographic sensors.
Answers are grounded in this paper and cite their sources.
May improve wearable PPG sensors in research; leaves open human validation before clinical consideration.
Wearable photoplethysmographic sensors in the form of wristbands and watches have become increasingly popular in mobile healthcare. We propose a technique to cancel motion artifacts by using two reflective pulse signals from a single green LED, switched by a microcontroller between high and low light intensities. We accurately estimate motion artifacts by applying differential measurement and common-mode rejection. The performance was assessed for four different motion artifacts: artificial white noise, artificial color noise, and vertical/horizontal random movements of a hand. Under each different condition, we accurately cancelled out motion artifacts.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2016) studied this question. Two reflective pulse signals from a single green LED switched between high and low light intensities was evaluated on Motion artifact cancellation under four different conditions. A technique using two reflective pulse signals from a single green LED switched between high and low light intensities accurately cancelled out motion artifacts under various conditions.
Synapse has enriched 10 related papers. Consider them for comparative context: